The Fluke 717 Pressure Calibrator is an essential tool for process technicians who demand precision and portability in the field. Designed to streamline the calibration and troubleshooting of pressure transmitters and loops, this instrument combines a highly accurate internal pressure sensor with a 24 mA current measurement and sourcing capability. Its intuitive interface features a clear 5-digit LCD, zero, min/max/hold, and damp functions, making it ideal for both quick spot-checks and detailed validation. With a rugged, compact body that withstands a 1-meter drop, the Fluke 717 is built for the harshest industrial environments, ensuring reliable performance year after year.
| Measurement Accuracy | ||
|---|---|---|
| Parameter | Fluke 717 Series | |
| Details | Specification | |
| Pressure (Internal Sensor) | Accuracy | 0.05% full scale |
| Functions | Zero, min, max, hold, damp | |
| Media Compatibility | Gases / liquids (non-corrosive) | |
| Range & Resolution | See table below for specific model ranges | |
| Model | Range | Resolution |
| Fluke 717 1G | -1 to 1 psi (-68.9 mbar to 68.9 mbar, -83 to 6.89 kPa) | 0.0001 psi, 0.001 mbar, 0.0001 kPa |
| Fluke 717 30G | -12 to 30 psi (-850 mbar to 2 bar, -83 to 206.84 kPa) | 0.001 psi, 0.1 mbar, 0.01 kPa |
| Fluke 717 100G | -12 to 100 psi (-850 mbar to 7 bar, -83 to 689.48 kPa) | 0.01 psi, 0.1 mbar, 0.01 kPa |
| Fluke 717 300G | -12 to 300 psi (-850 mbar to 20 bar, -83 to 2068.42 kPa) | 0.01 psi, 1 mbar, 0.1 kPa |
| Fluke 717 500G | 0 to 500 psi (0 mbar to 34.7 bar, 0 to 3447.38 kPa) | 0.01 psi, 1 mbar, 0.1 kPa |
| Fluke 717 1000G | 0 to 1000 psi (0 mbar to 69 bar, 0 to 6894.76 kPa) | 0.1 psi, 1 mbar, 0.1 kPa |
| Fluke 717 1500G | 0 to 1500 psi (0 bar to 103.4 bar, 0 to 10342 kPa) | 0.1 psi, 0.01 bar, 1 kPa |
| Fluke 717 3000G | 0 to 3000 psi (0 bar to 207 bar, 0 to 20684.27 kPa) | 0.1 psi, 0.01 bar, 1 kPa |
| Fluke 717 5000G | 0 to 5000 psi (0 bar to 345 bar, 0 to 34474 kPa) | 0.1 psi, 0.01 bar, 1 kPa |
| Fluke 717 1G | 5 x full scale | |
| All other G models (30G to 5000G) | 2 x full scale (except 300G: 375 psi / 25 bar) | |
| Measure Pressure | Via pressure module connector using optional Pressure Modules | |
| Range | 29 pressure modules, 0-1 inH2O to 10,000 psi, 2.5 mbar to 700 bar ranges. Over pressure per pressure module specs | |
| Resolution & Accuracy | Per pressure module specs | |
| Functions | Zero, min, max, hold, damp | |
| Current (mA) Measurement | ||
| Range | 0 to 24 mA (-25% to 125%) | |
| Resolution | 0.001 mA | |
| Accuracy | 0.015% + 2 counts | |
| Loop Power Supply | ||
| Voltage | 24 V DC | |
| Accuracy | ±10% | |
| Drive Capability | 20 mA into 1000 ohms for battery > 6.8 V; 700 ohms for battery 5.8 to 6.8 V | |
| Environmental Specifications | ||
| Operating Temperature | -10°C to 55°C | |
| Storage Temperature | -40°C to 60°C | |
| Humidity (without condensation) | 95% (10°C to 30°C) | |
| Operating Altitude | 3,000 m max | |
| Safety & Mechanical Specifications | ||
| Safety Compliance | CSA C22.2 No. 1010.1 1992 | |
| Dimensions (L x W x H) | 201 x 98 x 52 mm (7.93 x 3.86 x 2.06 in) | |
| Weight | 22 oz, approximately 600 g with holster | |
| Shock & Vibration | Random, 2 G, 5-500 Hz; 1 m drop test | |
| Battery Type | Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR619V alkaline) |
This is a family-level listing for the Fluke 717 Series, and no single pressure range is asserted for it. The filed sheet lists nine models: 717 1G, 717 30G, 717 100G, 717 300G, 717 500G, 717 1000G, 717 1500G, 717 3000G and 717 5000G, and each model's range is stated with its own model attribution.
717-1G: range (-1 to 1) PSI, (-7 to 7) kPa; max 5 PSI, 34.5 kPa. 717-30G: range (-12 to 30) PSI, (-83 to 207) kPa; max 60 PSI, 413 kPa.
717-100G: range (-12 to 100) PSI, (-83 to 690) kPa; max 200 PSI. 717-300G: range (-12 to 300) PSI; max 375 PSI. 717-500G: range (0 to 500) PSI; max 1000 PSI. 717-1000G: range (0-1000) PSI; max 2000 PSI. 717-1500G: range (0-1500) PSI; max 3000 PSI. 717-3000G: range (0-3000) PSI; max 6000 PSI. 717-5000G: range (0-5000) PSI; max 10000 PSI.
Pressure accuracy is +/- 0.05 % of range. The temperature coefficient is 0.01 % of range per °C for temperature ranges between -10 °C to 18 °C and 28 °C to 55 °C.
It measures current up to 24 mA, with a 24 mA range, 0.001 mA resolution and accuracy of 0.015 + 2 (± % of Reading + Counts), and fuseless overvoltage protection. It sources loop voltage to 24 V dc, and the loop supply is 24 V dc nominal.
Power requirements are a single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61). Safety compliance is IEC 61010-1 2nd Edition CAT I, 30 V; Pollution Degree 2.
Specifications follow the cited manufacturer documents and the approved model-specific conditions. Equipment mass is not a shipping-weight estimate.