The Fluke 725 Multifunction Process Calibrator is a versatile, high-accuracy tool designed for process control technicians who need to calibrate, troubleshoot, and verify a wide range of industrial instrumentation. This rugged calibrator combines sourcing and measuring capabilities for DC voltage, DC current, resistance, frequency, and temperature (RTDs and thermocouples) in a single, handheld device. Its dual display allows you to view input and output signals simultaneously, making it ideal for loop checks and transmitter calibration. With support for 13 thermocouple types and 7 RTD types, plus a built-in 24 V loop power supply, the Fluke 725 is a complete solution for maintaining process control systems.
| Measurement Accuracy | ||
|---|---|---|
| Voltage DC | 30.000 V | 0.02% + 2 counts (upper display) |
| 30.000 V | 0.02% + 2 counts (lower display) | |
| 100.00 mV | 0.02% + 2 counts | |
| -10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
| Current DC | 24.000 mA | 0.02% + 2 counts |
| Resistance | 0.0 to 400.0 Ω | 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire) |
| 401 to 1500 Ω | 0.5 Ω (4-wire), 1 Ω (2- and 3-wire) | |
| 1500 to 3200 Ω | 1 Ω (4-wire), 1.5 Ω (2- and 3-wire) | |
| Frequency | 2.0 to 1000.0 CPM | 0.05% + 1 count |
| 1.0 to 1100.0 Hz | 0.05% + 1 count | |
| 1.00 to 10.00 kHz | 0.05% + 1 count | |
| Sensitivity | 1 V peak-to-peak minimum | |
| Pressure | Accuracy from 0.025% of range using any of 29 pressure modules (modules available for differential, gauge, vacuum, absolute, dual, and high pressure) | |
| Source Accuracy | ||
| Voltage DC | 100.00 mV | 0.02% + 2 counts |
| 10.000 V | 0.02% + 2 counts | |
| -10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
| Current DC | 24.000 mA (source) | 0.02% + 2 counts |
| 24.000 mA (simulate) | 0.02% + 2 counts | |
| Resistance | 15.0 to 400.0 Ω | 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
| 401 to 1500 Ω | 0.5 Ω (excitation current 0.05 to 0.8 mA) | |
| 1500 to 3200 Ω | 1 Ω (excitation current 0.05 to 0.4 mA) | |
| Frequency | 2.0 to 1000.0 CPM | 0.05% |
| 1.0 to 1100.0 Hz | 0.05% | |
| 1.00 to 10.00 kHz | 0.25% | |
| Waveform | 5 V p-p squarewave, -0.1 V offset | |
| RTDs and Thermocouples | ||
| Measure accuracy | NI-120 | 0.2°C |
| PT-100 (385) | 0.33°C | |
| PT-100 (393) | 0.3°C | |
| PT-100 (JIS) | 0.3°C | |
| PT-200 (385) | 0.2°C | |
| PT-500 (385) | 0.3°C | |
| PT-1000 (385) | 0.2°C | |
| Resolution | 0.1°C | |
| J | 0.7°C | |
| K | 0.8°C | |
| T | 0.8°C | |
| E | 0.7°C | |
| R | 1.8°C | |
| S | 1.5°C | |
| B | 1.4°C | |
| L | 0.7°C | |
| U | 0.75°C | |
| N | 0.9°C | |
| Resolution | J, K, T, E, L, N, U: 0.1°C, 0.1°F; B, R, S: 1°C, 1°F | |
| XK | 0.6°C | |
| BP | 1.2°C | |
| Source accuracy | NI-120 | 0.2°C |
| PT-100 (385) | 0.33°C | |
| PT-100 (393) | 0.3°C | |
| PT-100 (JIS) | 0.3°C | |
| PT-200 (385) | 0.2°C | |
| PT-500 (385) | 0.3°C | |
| PT-1000 (385) | 0.2°C | |
| Resolution | 0.1°C | |
| Accuracy stated for 4-wire measurement. | ||
| J | 0.7°C | |
| K | 0.8°C | |
| T | 0.8°C | |
| E | 0.7°C | |
| R | 1.4°C | |
| S | 1.5°C | |
| B | 1.4°C | |
| L | 0.7°C | |
| U | 0.75°C | |
| N | 0.9°C | |
| Resolution | J, K, T, E, L, N, U: 0.1°C; B, R, S: 1°C | |
| XK | 0.6°C | |
| BP | 1.2°C | |
| Specifications | ||
| Ramp functions | Source functions | Voltage, current,
Fluke 725 Process CalibratorDoes the Fluke 725 have its own internal pressure range?No internal pressure range is stated for the 725. The manual states that pressure range, resolution and accuracy are determined by the pressure module, and the printed unit list includes psi, inH2O@4 °C, inH2O@20 °C, kPa, cmH2O@4 °C, cmH2O@20 °C, bar, mbar, kg/cm2, mmHg and inHg. Which thermocouple types can the 725 measure and source?The manual lists J, K, T, E, R, S, B, L, N, U, XK and BP for measure and source. Resolution is 0.1 °C and 0.1 °F for J, K, T, E, L, N, U, XK and BP, and 1 °C and 1 °F for B, R and S. What accuracy does the 725 specify for type K thermocouples?For type K, the manual gives 1.2 °C accuracy over -200 to 0 °C and 0.8 °C accuracy over 0 to 1370 °C. What does the 725's loop power supply provide?The loop power supply is 24 V with a maximum current of 22 mA and is short circuit protected. What are the 725's operating and storage temperature ranges?Operating temperature is -10 °C to 55 °C and storage temperature is -20 °C to 71 °C. What power source does the 725 use?The manual specifies power requirements of 4 AA alkaline batteries.
Specifications follow the cited manufacturer documents and the approved model-specific conditions. Equipment mass is not a shipping-weight estimate. Unmatched Customer ServiceAfter Hours ShippingReal Time Order TrackingFlexible Delivery OptionsIntact Technical SupportsNeed Professional Tech Support or a Quote?